MAY-2001T Input on-line density and concentration meter

QR code
MAYZUM MAY-2001T input on-line density meter is a continuous on-line measurement of liquid density instrument, mainly used in industrial process automation control. The corresponding 4-20mA or 485 analog signal is generated according to the media density change, and the digital signal can be converted to facilitate remote calibration and monitoring. Depending on the industrial process, direct reading is displayed: alcohol (Vol), Baume (° Be), plato (plato) and concentration percentage (%), etc. It can be widely used in industrial sites such as on-site pipelines, loose ports or closed storage tanks, and can continuously and real-time online measure the liquid density of open tanks or tanks.
type Input on-line density and concentration meter
brand 秒准MAYZUM
Model number MAY-2001T
Chen 18926057813
Online consultation
DetailDetailed parameterMatching spare parts

2001T.jpg


Product introduction:

MAYZUM MAY-2001T input on-line density meter is a continuous on-line measurement of liquid density instrument, mainly used in industrial process automation control. The corresponding 4-20mA or 485 analog signal is generated according to the media density change, and the digital signal can be converted to facilitate remote calibration and monitoring. Depending on the industrial process, direct reading is displayed: alcohol (Vol), Baume (° Be), plato (plato) and concentration percentage (%), etc. It can be widely used in industrial sites such as on-site pipelines, loose ports or closed storage tanks, and can continuously and real-time online measure the liquid density of open tanks or tanks.


Installation mode

2001T 安装方式.jpg


Technical parameter:

Product model

MAY-2001T

Density (concentration) accuracy

0.001 g/cm3 (0.19%).

Range of work

0 ~ 2.2 g/cm after. 0 ~ 100%

Output signal

4-20mA, RS485, switching quantity, etc

Process temperature

Standard: -25~+120, custom: -25C~+150

Ambient temperature

- 25 ~ + 85

Temperature coefficient

20ppm/°C(after correction)

Material for receiving liquid

ZSS316L, Hastelloy, mouth ceramic, PTFE Teflon

Working pressure

Standard within 2.5MPa, mouth special custom within 6.3MPa

Type of process connection

Input type

Probe size

Standard 130mm, can be customized 0 to 3 meters insertion depth

Explosion-proof certification

Ex dblIC T6 Gb

Power Requirements

DC24V,100mA

Electrical interface

M20 * 1.5

Viscosity range

0-2000cP(for higher viscosity choose MAY-5001S)




Product model

MAY-2001T

Density (concentration) accuracy

0.001 g/cm3 (0.19%).

Range of work

0 ~ 2.2 g/cm after. 0 ~ 100%

Output signal

4-20mA, RS485, switching quantity, etc

Process temperature

Standard: -25~+120, custom: -25C~+150

Ambient temperature

- 25 ~ + 85

Temperature coefficient

20ppm/°C(after correction)

Material for receiving liquid

ZSS316L, Hastelloy, mouth ceramic, PTFE Teflon

Working pressure

Standard within 2.5MPa, mouth special custom within 6.3MPa

Type of process connection

Input type

Probe size

Standard 130mm, can be customized 0 to 3 meters insertion depth

Explosion-proof certification

Ex dblIC T6 Gb

Power Requirements

DC24V,100mA

Electrical interface

M20 * 1.5

Viscosity range

0-2000cP(for higher viscosity choose MAY-5001S)


12 选购件(MAY-2001系列).png